Product Description
The system organically combines the pumps, pipelines, valves, bubble traps, ultraviolet detectors, etc. required by the conventional chromatography process, and has a complete pipeline system to ensure the stability and repeatability of the process. Customers can also choose air bubble detectors, conductivity detectors, flow meters, and increase or decrease material inlets and outlets according to their own needs. The system can also be customized according to user needs and special requirements of product technology.
Typical application of the system:
Monoclonal antibody field
Recombinant protein field
Vaccine field
Blood products field
Biochemical medicine field
Antibiotic field
Natural products field
CBD, THC field
Design Features:
The G3.3 glass column barrel is precision-processed by German SCHOTT, which has high dimensional accuracy and reliable sealing. And the inner wall has undergone secondary processing, without bubbles, scratches, and residues.
The O-ring sealing method is adopted between the column head and the column barrel, which is simple and safe.
Optimized design, no dead ends, no residues and minimum dead volume.
The unique distribution design can make the fluid fully diffuse.
The use of metal sieve plate makes it easier to discharge air bubbles.
Ergonomic structure design, simple operation, easy disassembly and maintenance of upper and lower sieve plates.
Model No |
HI Flow3L |
HI Flow10L |
HI Flow20L |
HI Flow40L |
HI Flow70L |
Flow range (L/h) |
1-180 |
6-600 |
60-1200 |
120~2400 |
120-4200 |
Operating pressure |
6 bar(40℃) |
6 bar(40℃) |
6 bar(40℃) |
6 bar(40℃) |
6 bar(40℃) |
Pressure Sensor |
0-10 bar ±0.1 |
0-10 bar ±0.1 |
0-10 bar ±0.1 |
0-10 bar ±0.1 |
0-10 bar ±0.1 |
Flow Sensors |
3/8” |
1/2” |
3/4” |
1.5” |
1.5” |
pH (optional) |
0-14①, ±0.02 |
||||
Conductivity range (optional) |
0.001-300ms/cm, ±5uS/cm |
||||
UV (optional) |
Single wavelength 280nm/variable four wavelength 200-400nm |
||||
Manual valve (optional) |
SS316L Ra ≤0.4µm PTFE/EPDM |
||||
3/8” |
1/2” |
3/4” |
1” |
1.5” |
|
Pump body |
SS316L:1.4435 |
||||
Check valve |
EPDM |
||||
Diaphragm |
TPE |
||||
Pipe fittings |
PP/SS316L Ra ≤0.4µm |
||||
3/8” |
1/2” |
3/4” |
1.5” |
1.5” |
|
Sanitary connection |
3/8”Triclamp |
1/2”Triclamp |
3/4”Triclamp |
1.5”Triclamp |
1.5”Triclamp |
Electrical parts |
SIEMENS S7 |
||||
Operation method |
HI Base software, automatic recording and analysis |
||||
Analyzing Software |
HIBase |
||||
Operating system |
Win10 |
||||
Pump cavity volume |
15ml |
75ml |
75ml |
820ml |
820ml |
Volume per revolution |
1.2ml |
5.8ml |
9.6ml |
52ml |
95ml |
Maximum speed |
3000rpm |
2000rpm |
2000rpm |
1200rpm |
1200rpm |
T-ring |
EPDM (USP Class VI) |
||||
Maximum temperature |
fluid |
80(℃) |
|||
CIP |
90(℃) |
||||
SIP |
130(℃) |
||||
Autoclave |
130(℃) |
||||
Power |
230VAC 0.5Kw |
230VAC 1.0Kw |
230VAC 1.0Kw |
400VAC 2.5Kw |
400VAC 2.5Kw |
Operating temperature |
4-40℃@PP / 4-60℃@SS316L ② |
||||
Size |
430/530/580mm |
430/530/580mm |
430/530/580mm |
/ |
/ |